The code ':' between the beat values has a special meaning, and can
also be '!' or '?' (see Table 4.4 Codes between the values of beat-to-
beat BP and HR).
Code
Description
:
Normal situation, good quality recording.
?
Beat values may have become dubious, full arterial contraction is
close.
!
Beat values are dubious due to arterial contraction.
Table 4.4 Codes between the values of beat-to-beat BP and HR
In Run mode when full arterial contraction is close, a typical display
looks like, e.g.:

4.2.5 Error mode

In Error mode Finometer® Model-2 has detected an error condition.
Recovery is through intervention of the operator.
A typical error message looks like:
This message indicates that an error occurred in the air supply (e.g.
pump unit, Frontend Unit air hose or finger cuff air hose). The
measurement is stopped permanently. Press the
that the message has been noticed. After pressing the
Finometer® Model-2 switches to Ready mode.
Systolic Diastolic Mean
127? 84?102? 78?
CONTRCTED ARTERY
Physiocal
Systolic Diastolic Mean
ERROR MODE: STOP
CHECK AIR SUPPLY
Physiocal
Finometer® Model-2
Rate (bpm)
Height
Rate (bpm)
Height
key to confirm
STOP
START
47