Stopping A Measurement; Data On A Pc - FINAPRES Finometer Model-2 User Manual

Table of Contents

Advertisement

68 Finometer® Model-2
T
he run time messages are defined in the following table (Table 6.2 Run
time messages).
Message
CONTR
CTED ARTERY
LOW PLETHYSM
Table 6.2 Run time messages
6.6
Stopping
Upon completion of the m
1.
Press the
START
2. Disconnect the cuff air hoses from the Frontend Unit to protect the
finger cuffs from accidental inflation.
3. Unwrap and remove the finger cuffs and take the Frontend Unit
away from the subject.
6.
7

Data on a PC

To start a Finometer® Model-2 recording on a PC:
1. Connect the serial cable between main unit and a serial port in the
PC.
2. Run the BeatScope® Easy PC software.
3. Refer to the BeatScope® Easy User's Guide for further information.
4. Make sure that the FMS serial proto
User's Guide
Systolic Diastolic Mean
SSS!DDD!MMM!RRR!
LOW PLETHYSM
Physiocal
Description
No plethysmogram i
hands:
Warm the hand and the arm proximal of the cuff.
Try another finger.
Pulsatile component (V:) of plethysmogram is too
small:
• Warm the hand and t
• Try another finger.
a measurement
easurement, it is stopp
key to stop the measurement.
/
STOP
Rate (bpm)
Height
s observed, probably due to cold
he arm proximal of the cuff.
ed as follows:
col is configured, see 5.4.4.