How to optimize the detector

Set the Detector Parameters

Set the Detector Parameters
1 Set peakwidth as close as possible to the width (at half height) of a
2 Choose the sample wavelength
3 Consider to use time- programming to further optimization.
78
narrow peak of interest. Refer to
at a longer wavelength than the cut- off wavelength of the mobile
phase,
at a wavelength where the analytes have strong absorptivity if you
want to get the lowest possible detection limit,
at a wavelength with moderate absorptivity if you work with high
concentrations, and
preferably where the spectrum is flat for better linearity.
"Peakwidth Settings"
Agilent 1260 Infinity VWD User Manual
on page 71.