Cover Violation

Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If
the foam is removed while the lamps are on (or if an attempt is made to
switch on for example the lamps with the foam removed), the lamps are
switched off, and the error message is generated.
Probable cause
1
2
Agilent 1260 Infinity VWD User Manual
The top foam was removed during
operation.
Foam not activating the sensor.
Error Information
Detector Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.