Guidance and manufacturer´s declaration – electromagnetic emissions/
immunity
The plusoptiX 09 is intended for use in the electromagnetic environment specified below. The
customer or the user of the equipment should assure that it is in such an environment.
Emissions Test
RF Emission CISPR 11
RF Emission CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC
61000-3-3
Immunity test
ESD IEC 61000-4-2
Electric
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, short interruptions and
voltage variations on power supply
input lines IEC 61000-4-11
Power Frequency (50/60 Hz) magnetic
field IEC 61000-4-8
Compliance
Electromagnetic environment - guidance
The plusoptiX 09 uses RF Energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
Class B
The plusoptiX 09 system is suitable for use in all establishments, including
Class B
domestic establishments and those directly connected to the public low
Voltage power supply network that supplies buildings used for domestic
purposes.
Complies
IEC 60601 test level
Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power supply
± 2 kV for power supply
lines
lines
± 1 kV differential mode
± 1 kV differential mode
± 2 kV common mode
± 2 kV common mode
0%
0.5
periods
0°
0%
40%
5
periods
0°
40%
70%
25
periods
0°
70%
0% 250 periods 0°
0% 250 periods 0°
3 A/m
3 A/m
Page 33 of 34
Electromagnetic environment –
Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material the relative humidity should be at
least 30 %
Mains power supply should be that of a
typical commercial or hospital environment
Mains power supply should be that of a
typical commercial or hospital environment
0.5
periods
0°
5
periods
0°
Mains power supply should be that of a
25
periods
0°
typical commercial or hospital environment
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.