Operation
5.4 Operating functions
5.4.3

Calibration

Note
The calibration tolerance (function 78) is permanently active in order to monitor for an
incorrect calibration or leaks in the system during the calibration procedures.
5.4.3.1

Calibration submenu

Calibration
2O Zero calibration
21 Span calibration
22 Setpoints for zero/
23 Calibration settings
Figure 5-8
You can calibrate the device using the functions available in this menu. Press the fifth
softkey "...continue" to access further functions.
This menu is protected by the code of code level 1.
The OXYMAT 64 offers the option of a manual or automatic calibration (Autocal:
function 24). The latter is only possible with an optional board which contains an additional
eight binary outputs and eight relay outputs.
The setpoints for the zero and span calibrations are entered using function 22.
Note
If the analyzer has been equipped with an optional board with Autocal function, it
automatically enters Autocal mode when switched on. In order to carry out a manual
adjustment or via binary inputs, the Autocal mode must first be switched off (function 24).