5.4.4
Measuring ranges
5.4.4.1
Measuring ranges submenu
Meas.ranges
4O Range selection
41 Define meas. ranges
Figure 5-25
The measuring ranges menu contains all functions you need for selecting and setting the
measuring ranges.
This menu is protected by the code of code level 1.
5.4.4.2
Select measuring ranges (function 40)
4O Select ranges
MR1
MR2
MR3
MR3
Auto range
Figure 5-26
You can select a fixed measuring range or switch to automatic measuring range switching.
All selection options are subject to mutual interlocking.
Automatic measuring range switching is only possible under the following conditions:
● At least two measuring ranges must be available. One measuring range is said to be
● The spans must become greater or smaller
● The measuring ranges must "border on" each other or overlap
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
Measuring ranges submenu
O.O -
1O.O vpm
O.O - 1OO.O vpm
O.O - 1OOO.O vpm
O.O - 1OOOO.O vpm
Select measuring ranges
present when the following is true: Measuring range start-of-scale value ≠ full-scale value
O
2
O
2
Operation
5.4 Operating functions
67