Siemens OXYMAT 64 Operating Instructions Manual page 96

Gas analyzer for measurements of trace oxygen
Hide thumbs Also See for OXYMAT 64:
Table of Contents

Advertisement

Operation
5.4 Operating functions
Relay and binary test
Relay and Binary
No
1
2
3
4
5
6
Figure 5-60
NOTICE
The data connector must be removed before calling this function!
The first function screen shows the state of the six relay and binary inputs of the standard
device. With an optional board, there are another eight relay/binary inputs on a second page.
With the relay test, individual relays can be activated. This is done via the input field. With
"1", the relay is on, with "0", it returns to its release condition. Numbers other than "0" and "1"
are not accepted by the input field.
After leaving function 80, the relays are back in the state they had before calling the relay
and binary test.
The current state of the binary inputs is displayed in the "Binary" column.
Analog test
With the analog test, the analog output is set to a constant current of 0 to 24000 µA for
testing purposes.
The analog input displays the input currents in µA.
94
Relay
Binary
:1:
O
1
O
0
O
0
O
1
O
0
O
...continue
Test relay outputs and binary inputs
O
2
Operating Instructions, 06/2008, A5E00880383-03
OXYMAT 64