Suppression of short-term noise signals (function 76)

76 Suppress fault..
Suppress noise signals
with a duration of up to
Level in % of
smallest MR:
Figure 5-55
With function 76, you can eliminate undesired spikes which exceed a settable threshold.
Spikes are caused by electromagnetic interferences or occasional mechanical shocks.
These faults can be suppressed by entering a "blanking time" of 0 to 5 s. The entered time
has the effect that the last measured value before the spike occurred is displayed so that the
measured result is not influenced.
Times can be entered in steps of 0.1 s.
Under
range above which the noise signals are to be suppressed.
Note
If a change in concentration directly follows a fault, this may be displayed after a delay.
When this function is activated, the settings of function 50 ("Electric time constants") must be
taken into account, since these are executed first.
See also
Electric time constants (function 50) (Page 71)
90
: O.O: s
12.O %
Suppression of noise signals
"Level in % of smallest MR"
O
2
, enter the threshold value in % of the smallest measuring
Operating Instructions, 06/2008, A5E00880383-03