4

Preparing Solvents

Proper solvent selection and preparation are critical in differential refractometry,
to prevent an erratic baseline or effects such as drift and noise.

4.1 Common solvent problems

The 2414 RI detector measures changes in the refractive index (RI) of the solution
flowing through the sample side of the flow cell. However, factors other than the
presence of dissolved sample molecules can affect a solution's RI. Common problems
include these conditions:
Changes in temperature
Changes in pressure
Contaminants
Separation of mixed solvents
Outgassing of dissolved gases

4.2 Selecting a solvent

An ideal solvent for your analysis has these characteristics:
Good solubility characteristics for your application
Significantly different RI than that of the sample components
Satisfactory baseline noise performance
Optimum optical sensitivity characteristics
Warning:
To avoid the harmful effects of personal contact
with solvents, including inhalation, always wear
chemical-resistant, powder-free gloves and observe good
laboratory practice when you handle them. See the Material
Safety Data sheets for the solvents you use.
Warning:
To avoid the harmful effects of personal contact
with leaking solvents, flush the detector with an appropriate
miscible solvent prior to installing or replacing fittings.
June 11, 2015, 715004754 Rev. A
Page 57