Troubleshooting - Agilent Technologies 1220 Infinity II LC Series User Manual

Sim sample cooler
Table of Contents

Advertisement

8. TROUBLESHOOTING

The error (see Tab. 3) occurs, because the Peltier element becomes too warm/hot:
 When the Peltier element is cooled down, the cooler restarts automatically.
Tab. 3
Troubleshooting
Error
Temperature display
becomes orange
(Warning, in log-data is
displayed: "Peltier hot side
is at warning level.")
Temperature display gets
red and Peltier element is
switched off
(in log-data is displayed:
"Peltier hot side is at error
level; disabled Peltier.")
28
Probable Cause
• no cooling water
(see viewing window)
• fan is covered or
damaged
• ambient temperature is
too high
• cooling water is not
pumped over any more
(check the return flow in
the water tank!)
Scientific Instruments Manufacturer GmbH
D-Oberhausen ● +49 (0)208/ 94 10 78-0 ● www.sim-gmbh.de
Suggested Action
• contact service engineer
• check the distance to other
devices / the wall (minimum
distance 10 cm), clean the fan
grill or contact the service
engineer
• make sure that the ambient
temperature is in the claimed
range (5 – 35 °C)
• remove the cover of the
cooling module and check if
the cooling hoses are kinked
• contact service engineer
User Manual SIM Sample Cooler_06/2018