Opt024 Driver For Openlab Cds; Site Requirements - Agilent Technologies 1220 Infinity II LC Series User Manual

Sim sample cooler
Table of Contents

Advertisement

2.5. OPT024 Driver for OpenLAB CDS 2.x
 Software (LC 1220 10 24)
 2 LAN cables (0.25 m and 2.0 m, LC 1220 10 75)

3. SITE REQUIREMENTS

DANGER/ DANGER
Absence of ground connection or use of unspecified power cord!
Absence de raccordement à la terre et utilisation d'un câble d'alimentation
non adapté!
The absence of ground connection or the use of an unspecified power cord
can lead to electric shock or short circuit.
L'absence de mise à la terre ou l'utilisation d'un câble d'alimentation non
recommandé peuvent entraîner des chocs électriques ou des courts-
circuits.
Never connect the device to a voltage source without ground
connection.
N'utilisez jamais votre instrument avec une prise de courant sans mise à la
terre.
Never use cables other than the ones supplied by SIM and Agilent
Technologies to ensure proper functionality and compliance with safety or
EMS regulations.
N'utilisez jamais de câble d'alimentation autre que le modèle SIM et Agilent
Technologies destiné à votre pays.
NOTE
In case of emergency it must be possible to disconnect the instrument from the power line
any time.
Make sure that there is easy access to the power cable of the instrument so that the
instrument can quickly and easily be disconnected from the line voltage.
Power considerations
Power supply:
The power supply has wide ranging capabilities so that there is no voltage selector at the
instrument. The fuse (T 3.15 A / 250 VAC) is accessible from the outside.
Power cord:
Different power cords are offered as options with the system. The female ends of all power
cords are identical. The female end plugs into the power-input socket at the rear of the
cooling module. The male end of each power cord is different and designed to match the wall
socket of a particular country or region.
User Manual SIM Sample Cooler_06/2018
Scientific Instruments Manufacturer GmbH
D-Oberhausen ● +49 (0)208/ 94 10 78-0 ● www.sim-gmbh.de