Step 5
Initial Solvent: The volume of the first solvent used is then
selected from between 0 mL and 500 mL.
NOTE 1: If too little solvent is used for the extraction pouch size
then an error screen is shown when Run is selected.
The 100 mL extraction pouch requires a minimum of 25 mL of
solvent to ensure good mixing, the minimum amount required
for the 500mL extraction pouch is 100 mL.
NOTE 2: If too much solvent is used for the 100 mL extraction
pouch then another error screen is shown when Run is
selected.
The maximum amount of solvent that can be used in this
extraction pouch is 100 mL.
46-208-208 Rev. B—XTRX500 User Guide