Configuration Menu Structures And Parameter Descriptions - Rice Lake 480 Legend Series Technical Manual

Digital weight indicator
Hide thumbs Also See for 480 Legend Series:
Table of Contents

Advertisement

3.2.1

Configuration Menu Structures and Parameter Descriptions

CONFIG
FORMAT
GRADS
ZTRKBN
010000
0 dd
DIGFIL
ADPFIL
ROLFIL
Parameter
Choices
GRADS
10000
1–100000
ZTRKBN
0 dd
0.0–3.0dd
2%
ZRANGE
1%–100%
INIZR
0%
0–100%
MOTBAN
0.1d
0.0d–100d
OVRLOA
FS+2%
FS+1D
FS+9D
FS
CALIBR
PROGRM
DIG IN
ZRANGE
INIZR
2 %
0 %
ADPFIL
See
Figure 3-10
on page 37
OFF
Figure 3-9. Configuration Menu
Graduations – specifies the number of full scale graduations; the value entered must be in the range
1–100,000 and should be consistent with legal requirements and environmental limits on system resolution
Zero track band – automatically zeros the scale when within the range specified, as long as the input is within
the configured zero range (ZRANGE parameter); selections are ± display divisions; maximum legal value
varies depending on local regulations
Zero range – selects the range within which the scale can be zeroed; the 2% selection is ± 2% around the
calibrated zero point, for a total range of 4%; indicator must be at standstill to zero the scale
Initial zero range at power up; See Error messages HINOFF and LINOFF in
power up, the weight value is between the ±% range specified of Calibrated Zero, the indicator will zero off
that weight; if outside that range, the indicator will display HINOFF or LINOFF, depending on whether the
initial reading is above or below the range
NOTE: If not set at zero, hopper/scale MUST be empty when powered up.
Motion band – sets the level, in display divisions, at which scale motion is detected; if motion is not detected,
the standstill symbol lights; some operations, including print, tare and zero, require the scale to be at standstill;
maximum legal value varies depending on local regulations
NOTE: Setting the MOTBAN setting to 0.0 disables motion sensing and the scale will always be
considered stable.
Overload – determines the point at which the display blanks and an out-of-range error message is displayed;
maximum legal value varies depending on local regulations
Table 3-10. Configuration Menu Parameters
ALGOUT
PASWRD
MOTBAN
OVRLOA
0.1 d
FS-2%
FS-1D
FS-9D
FS
ROLFIL
TAREFN
See
Figure 3-10
BOTH
on page 37
NOTARE
PBTARE
KEYED
Description
© Rice Lake Weighing Systems ● All Rights Reserved
Configuration
USBMEM
TEST
MENU
DSPRAT
SMPRAT
0.10 SEC
5 HZ
0.25 SEC
10 HZ
0.50 SEC
20 HZ
0.75 SEC
40 HZ
1.00 SEC
NONE
Section 9.1.1 on page
61; if on
35