4.3 Self-check

4.3.1 Self-check Procedure

To run the self-check of the emulator, do so as explained here below. While the self-check is in progress, the LEDs will
change as shown in Figure 4.3.
(1) If the user system is connected, disconnect it.
(2) Within 2 seconds of activating power to the emulator, press the system reset switch on the emulator front panel to
switch the emulator to maintenance mode.
(3) Check the SAFE LED starts flashing and then press the system reset switch again.
(4) The self-check will start. If the normal result is displayed in about 30 seconds, the self-check terminated normally.
POWER
POWER
SAFE
CLOCK
ERROR
RESET
RUN
WARNING
SYSTEM
TARGET
STATUS
STATUS
Maintenance mode
started
SAFE and ERROR flash
SAFE flashes and
ERROR lights
SAFE and ERROR
light alternately
SAFE goes out and
ERROR lights
Figure 4.3 LED displays in the self-check
POWER
SAFE
ERROR
SYSTEM
STATUS
Self-check terminated
normally
POWER
SAFE
ERROR
ERROR 1
SYSTEM
STATUS
Emulation probe and firmware
do not match
POWER
SAFE
ERROR
ERROR 2
SYSTEM
STATUS
PC7501 system status error
POWER
SAFE
ERROR
ERROR 3
SYSTEM
STATUS
Emulation probe initialization
POWER
SAFE
ERROR
ERROR 4
SYSTEM
STATUS
Emulation probe status error
: ON
POWER
: OFF
CLOCK
: Flashing
RESET
RUN
WARNING
TARGET
STATUS
POWER
CLOCK
・When the firmware does not support the emulation probe
RESET
・When the emulation probe is not connected properly
RUN
WARNING
TARGET
STATUS
POWER
CLOCK
・When the PC7501 system is not working properly
RESET
RUN
WARNING
TARGET
STATUS
POWER
CLOCK
・When a clock is not supplied to the emulation probe
RESET
RUN
WARNING
TARGET
STATUS
error
POWER
CLOCK
・When the emulation probe is not working properly
RESET
RUN
WARNING
TARGET
STATUS
( 41 / 58 )