Carousel Motor Board; 24V Dc Board; Pressure Transducer Board; General Troubleshooting - Teledyne Tekmar Versa User Manual

Table of Contents

Advertisement

Carousel motor board

The Carousel Board is located on the carousel mechanism
and can be seen from the front of the instrument. The
Carousel Board is responsible for for the carousel motor,
optical sensor and othe boards. The switch setting should
be set to position "one. "

24V dC board

The 24V DC Board is located on the bottom of the chassis
on the left-hand side. The 24V DC Board is responsible for
power supply to the valco valve, GC/IO board, elevator
board and carousel board.
5V dC board
The 5V DC Board is located on the bottom of the chassis
on the left-hand side. The 5V DC Board is responsible for
power supply to the temperature board and master board.

Pressure transducer board

The Pressure Transducer Board is located in the top
section of the chassis on the left-hand side. The Pressure
Transducer Board controls the communication between
the pressure transducer and Versa TekLink™2G.

generaL troUbLeshooting

Low response Loop

Perform a Leak Check on system
If the leak check fails, retry and cap Vent Bulkhead Fitting.
This will check for a leak or failure of the Vent Valve. Also,
try another non-punctured vial and repeat the Leak Check.
Versa User Manual
Maintenance & Troubleshooting • Page 7-9
Figure 7-15: Carousel Motor Board
Figure 7-16: 24V DC Board
Figure 7-17: 5V DC Board
Figure 7-18: Pressure Transducer Board
www.teledynetekmar.com