Therefore, the three variables, temperature, pressure and volume, have a direct affect on the sensitivity of the system.
The temperature in the Versa that would affect injection mass is the Valve Oven Temperature. The lower the temperature
used, the denser the mass of the gas in the loop. The pressure affects the sensitivity in two areas: the higher the vial
pressurization setting, the more dilution gas is added thus lowering sensitivity; while a higher loop fill pressure results in
higher sensitivity. The volume of the loop can be changed to allow for more mass to be injected on column. The volume
can be varied between the optional loop sizes of 100μl up to 5mL. The Versa is shipped with a 1mL loop.
Note: Refer to the Method Development section of this manual for further information on the effects of temperature,
pressure and volume.

Injection

Once the Sample Loop has been filled, and the GC is ready for injection and the 6-port valve is rotated. When the 6-port
valve is in the "B" position, it allows carrier gas to back flush the sample from the loop onto the analytical column. The GC
data system is started concurrently. The vial is lowered from the Platen and the sample pathway is subsequently swept
while waiting on the equilibration of the next sample to eliminate cross contamination.

system specIfIcatIons

Autosampler Capacity
Vial Size
Platen Heater
Sample Loop
Sample Mixing
Sample Pathway
Software
GC Interface
Pressure Control
Valving
Voltage Requirements
Gas Requirements
Unit Dimensions
Weight
Environmental
Conditions
Corrosion Resistance
Certifications
Electronic Leak Check
Benchmark Test
Diagnostics
E-mail Alert
Warranty
Versa User Manual
technical specifications
20 Positions
22 mL vials
Single position; ambient to 200° C via resistance heater settable in 1° C increments
Ships with a standard 1mL Silco® coated loop. Other optional loop sizes include 100, 250, 500µL as
well as 2, 3, and 5mL
System allows variable power settings from low/medium/high
Siltek® tubing. Entire sample pathway temperature controlled up to 200° C
TekLink™2G software in a Windows® XP or greater environment via USB connnection. TekLink™2G can
be configured for full 21 CFR Part 11 compliance tool capability
Interfaces to virtually all commercially available GC instruments. Supplies or accepts GC and Data
System start/ready signals via software selectable GC I/O configuration
Software has the ability to control pressure to assure consistent volume regardless of external
conditions for all samples via a pressure transducer
Motor actuated 6-port valve with removable rotor. Temperature controlled from ambient to 200° C.
100/115 VAC +/- 5%, 50/60 Hz, 5 amps & 220/240 VAC +/- 5%, 50/60 Hz, 2.5 amps
Ultra-high purity (99.999%) Helium or Nitrogen; incoming pressure: 45-140 psi
21" (53.34cm) H x 12" (30.48cm) W x 21" (53.34cm) D
40lbs (18.14kg)
Operating Temperature: 10° to 30°C; Storage Temperature; -20° to 60° C;
Relative Humidity: 10% to 90%
The chassis and carousel tray are resistant to aqueous samples with a pH range of 1-10
CE, CETL, CSA, ETL
Ability to leak check the sample pathway of the system via an automated system leak check process.
The system has a mode that will allow for full electromechanical testing including; valving, heater, vial
handling systems, liquid delivery system, inputs and outputs.
The system offers independent control of all valves and vial handling mechanisms and for troubleshooting.
The system can be configured to send an E-mail to alert the user of schedule completion or stoppage.
1 year from installation. 18 months from shipment.
service
Introduction • Page 1-7