Co2 Carrier Pcb Connector Pin Assignment - Dräger Evita 4 Repair Instructions

Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

5.2
CO
Carrier PCB Connector Pin Assignment
2
Connectors X1 and X2 are not considered in the following tables.
Pin
Assignment
1
+5 V-ISO-1
2
GND-ISO-1
3
GND-ISO-1
4
+15 V-ISO
5
GND-ISO-1
6
–15 V-ISO
7
–15 V-ISO
Pin
Assignment
1
GND-EXT
2
ECG-SYNC
Pin
Assignment
1
GND-ISO-1
2
5 V-ISO-1
Pin
Assignment
1
Temperature sensor
5664.500/800 Evita 4/Evita 2 dura
Dräger Medizintechnik
X3
Pin
8
9
10
11
12
13
14
X4
Pin
3
X5
Pin
3
4
X8
Pin
06/98
D
Assignment
TXD-SpO
2
RXD-SpO
2
CTS-SpO
2
Reset SpO
2
GND-ISO-1
ECG-SYNC (ISO)
GND-ISO-1
Assignment
Assignment
+15 V-ISO
–15 V-ISO1
Assignment
Page 35
Repair Instructions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evita 2 dura

Table of Contents