10.3
Replacing the Graphics Controller PCB
• Switch the Evita 4 off.
• Open the control unit.
• Remove the plug-in connection to the electronics.
• Remove the two hexagon head cap screws at the rear of the control unit.
Fig. 29: Rear view of the control unit – dismounting the Graphics Controller PCB
• Remove the six Phillips screws on the Graphics Controller PCB.
1
Fig. 30: Soldering side of the Graphics Controller PCB – dismounting the board
5664.500/800 Evita 4/Evita 2 dura
1
1
X4
X14
X2
06/98
Dräger Medizintechnik
1
X7
X1
8
1
Repair Instructions
D
Page 57