Conversions; Replacing Co2 Carrier Pcb According To Idm No. 3 (Evita 4) - Dräger Evita 4 Repair Instructions

Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

5.5

Conversions

5.5.1
Replacing CO
• Switch unit off.
• Observe electrostatic precautions. Use an ESD protection mat and a wrist strap.
• Dismantle CO
Carrier PCB (lowest PCB in the electronics unit).
2
• Replace resistors R82 and R84 with 511 ohms by 332 ohm resistors (order no.
18 16 144).
If the resistance of R82 and R84 cannot be identified by the serial number, you can
determin the ohmic resistance using a multimeter. It is also possible to determine the
ohmic resistance by the colour code. You have to replace the resistors R82 and R84
with the code "green – brown – brown – black – black" or
"green – brown – brown – black – red".
• Mark the PCB to be replaced with "IDM 3" on the back.
• Assemble the unit ready for operation.
• Switch unit on and activate CO
• Carry out function check in standby mode after 3 minutes.
• Switch to ventilation and call up CO
• Test plausibility of CO
the cuvette. Alternatively, test CO
Test Certificate.
• Check electrical safety following test items 10 to 10.3.1 of the Test Certificate.
5664.500/800 Evita 4/Evita 2 dura
Carrier PCB According to IDM No. 3 (Evita 4)
2
R 84
measurement via "alarm limits" and "monitoring".
2
real-time curve.
2
real-time curve and etCO
2
measurement according to test items 9 to 9.2 of the
2
06/98
Dräger Medizintechnik
measured value by breathing through
2
Repair Instructions
D
Page 39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Evita 2 dura

Table of Contents