Sub D Socket Configuration - Dräger Evita 4 Repair Instructions

Hide thumbs Also See for Evita 4:
Table of Contents

Advertisement

1.1.1

Sub D Socket Configuration

Pin
Assignment
1
+24 V –1680 mV +720 mV
2
+15 V ±450 mV
3
AGND
4
DGND
Important: The voltages can be read out in diagnosis mode
Instructions - Service No./ Modi, Error list" chapter 3.3.2, Diagnosis page "Microprocessor"
of "Electronics", page 70.
Instructions - Service No./ Modi, Error list" chapter 9.4.1, Testing voltages, page 138.
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.4.2, Power
pack status, page 140.
5664.500/800 Evita 4/Evita 2 dura
Pin
5
7
9
10
or with external DS mode using PC
06/98
Dräger Medizintechnik
Assignment
+5 V, 15 V ±77.3 mV
Shield
+12 V ±240 mV
–15 V ±450 mV
refer to section of "Repair
refer to section of "Repair
Repair Instructions