2

CPU 68332 PCB

CAUTION
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.
2.1

Power Supply to the CPU 68332 PCB

Pin
Assignment
1c
2c
3c
4c
Pin
1
3
Pin
1
3
Pin
32
16
5664.500/800 Evita 4/Evita 2 dura
X2 (Power supply only)
+12 V
+24 V
+15 V
–15 V
X11
X13
VCC RAM
+5 V (minimum voltage for retention of data: 2 V)
06/98
Dräger Medizintechnik
Pin
Assignment
8c
AGND
15c
DGND
16c
+5 V (also at X11)
Assignment
+5 V supply voltage
DGND
Assignment
+5 V ISO
GND ISO
Assignment
DGND