• Remove the empty housing.
• Disconnect the connectors D (on 2 single black cables coming from the mains filter)
from the power supply unit basic board and connect them to the battery module.
D
Fig. 9: Connections between the power supply unit and the DC module (part 1)
• Connect the 2-core cable E (with 3-pin plug housing) to the battery module. The
connector of the cable must be plugged onto the pins of the battery module such that
the retaining clamp engages firmly (the retaining clamp is located above the cut pin).
The 2-core cable E is located on older power supply units at the mains filter and is
secured with a cable tie. With newer power supply units the calbe is "parked" on the
PCB in the power suppy unit on a connector.
5664.500/800 Evita 4/Evita 2 dura
Dräger Medizintechnik
Mains
filter
06/98
D
Page 12
Repair Instructions