IMPORTANT
Dust particles may affect the view on the display. Before mounting the display,
clean the display and the window with a lint-free cloth.
CAUTION
Alcoholic cleaning agents may damage the display and the window. DO NOT
use alcoholic cleaning agents to clean the display or the window.
• Mount the window by following the reverse method used for dismounting.
• Check the Evita 4 by following the instructions in the Evita 4 Test Certificate.
5664.500/800 Evita 4/Evita 2 dura
Dräger Medizintechnik
06/98
D
Page 70