Table 8 - Rtd Constants - Newport 8000 User Manual

Modular controller
Table of Contents

Advertisement

Chapter 4
4.5.3.1
RTD Constants
The constants entered for an RTD depend on the type of curve it has. Table 8 shows
three standard types.
Curve
Laboratory
US
European
The Ro constant also applies for RTD sensors. It is nominally 100.00 Ω, but can be
varied from 95.00 Ω to 105.00 Ω.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Temperature Controller Module Operation
t is the temperature in °C.
TCR
C1
(Ω Ω Ω Ω /Ω Ω Ω Ω /° ° ° ° C)
.003926
3.9848x10
.003910
3.9692x10
.003850
3.9080x10

Table 8 - RTD Constants

C2
-3
-6
-0.58700x10
4.0000x10
-3
-6
-0.58495x10
-4.2325x10
-3
-6
-0.58019x10
-4.2735x10
C3
R0
-12
100.00
-12
100.00
-12
100.00
63

Advertisement

Table of Contents
loading

This manual is also suitable for:

8008

Table of Contents