46
Chapter 4
1.
Turn the 8000 power off. Installing a module with the 8000 on can damage the
module and the 8000.
2.
Place the module into an open bay on the back of the 8000 and slide the module
into place. There are tracks at the top and bottom of the bay which guide the
module into place. Screw the Module Locking Screws into the back panel to
secure the module.
To remove the module from the 8000 follow these steps:
1.
Turn the 8000 power off. Removing a module with the 8000 on can damage the
module and the 8000.
2.
Unscrew the Module Locking Screws which secure the module to the 8000 back
panel.
3.
Grasp the module by handle which extends from the bottom of the back panel.
Gently, but firmly, pull the module out of the 8000.
4.2

TEC Safety Features

4.2.1
Conditions Which Will Automatically Shut Off the TEC
Output
High Temperature Limit
Low Temperature Limit
R Limit
Sensor Open
TEC Module Open
Sensor Select changed
Sensor Shorted
Any Linked Functions
With the exception of the linked functions, each of these conditions can be disabled
by clearing the appropriate bits in the TEC OUTOFF register. See the Computer
Interfacing Manual for additional information.
4.3

The TEC Connectors

On the TEC Module, a 15-pin D-connector is used for input and output connections
to the series 700 mounts, as shown by the pin-out diagram below.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Temperature Controller Module Operation