Chapter 3
3.
On all 700 series mounts, the temperature control cable must be connected for
maximum protection.
4.
For maximum laser diode protection, the diode should be recessed, as in the 700-
C mount.
3.2

Laser Safety Features

3.2.1
Conditions Which Can Automatically Shut Off the Laser
Output.
With the exception of the linked functions, some of these conditions can be disabled
by clearing the appropriate bits in the Laser OUTOFF register. See the Computer
Interfacing Manual.
3.2.2

Key switch Interlock

The LASER ENABLE key switch on the front panel will shutoff, or not allow to be
turned on, any laser outputs while in the OFF position, per CDRH requirements.
3.2.3

Turn On Delay

The 8000 is CDRH Compliant with a user programmable turn on delay. The default
turn on delay is three seconds, but is user programmable from 0 to 30 seconds. The
delay setting is in the system configure screen, which can be reached by pressing the
MENU button, followed by the Config soft key, then the System soft key. The field
is labeled On Delay.
3.3

The Laser Connectors

On all laser modules, except the MOPA, a 9-pin female D-connector is used for input
and output connections, as shown by the pin-out diagram below.
1
This condition will always shutdown the laser output, and cannot be disabled.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Laser Diode Driver Module Operation
Laser Open Circuit
Laser Compliance Voltage Limit
Laser Hard Current Limit
Laser Interlock State Changed
Laser Photodiode High Current Limit
Laser Photodiode High Power Limit
Laser Intermittent Contact (if enabled, default disabled)
A Linked Function
Pin
Description