Chapter 6
6.3

TEC Calibration

This chapter describes how to calibrate the 8300/8600 Series modules.
6.3.1

Recommended Equipment

Recommended test equipment for calibrating the module is listed in Table 11.
Equipment other than that shown in the table may be used if the specifications meet
or exceed those listed.
Description
DMM
Resistors
Resistor
Connector
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Calibration
g.
Input the second actual (measured) voltage as in Step e.
h.
Once the actual voltage value is entered via the LAS:LDV command,
the 8000 leaves the current calibration mode.
If, at any time prior to the second LAS:LDV, a command other than
LAS:LDV or LAS:CAL:LDV? is sent to the 8000, the 8000 will
cancel the calibration mode and then process the command(s). The
*OPC? query may be used to determine when the calibration is
completed.
The operation complete flag (bit 0 of the Standard Event Status
Register) may be used to trigger an interrupt. This type of interrupt is
enabled by setting bit 0 of the Service Request Enable register and
using the *OPC command.
Mfg./Model
HP34401A
Metal Film
High Power
D-sub

Table 11 - Recommended Test Equipment

Specification
DC Amps @ 1.0 A): ±1%
Resistance (@ 10 ohms): 0.02%
10 kΩ for I
calibration (dual/combo
TE
only)
20 kΩ for 100µA calibration
200 kΩ for 10µA calibration
3 kΩ for LM335 sensor calibration
16 kΩ for AD590 sensor calibration
100 Ω for RTD sensor calibration
1 Ω, 50 W, for current calibration
15-pin male