This will prevent having the contents of the second vial being injected
into both position 1 and position 2. If the SampleList is already at the end
of the selected range, the second injection is not performed and the
message ' No more vials in selected range' displayed in the message
line. If No is selected, the AutoSampler will use the contents of one vial
to inject into each injector. The default value is Yes.
Clean Between Injections? (Yes or No) – This parameter is only used if
the Injector Position is set to Both. Setting this parameter to Yes will
cause the CP-8400 to execute the clean steps specified in the Clean
Mode softkey of the method. Setting this parameter to No will cause the
system to execute the second injection without cleaning. The default
value is No.

Priority Sample

Priority Sample functions to allow you to interrupt automation to run a
sample that needs to be run as soon as possible. After the priority
sample(s) have been completed the automation that had been running
will resume where it left off. Please note that Prep-Ahead is not available
for Priority Samples. Any time that the CP-3800 is running one of its
automation modes, the Priority Sample will be available; and can be
initiated at any time using the Run Now softkey. The current run in
progress will continue to its conclusion before the Priority Sample is run.
The system will return to the next replicate of the run that was in
progress or continue with the next scheduled injection once the Priority
Sample is completed. The numbering of the carrousel for the CP-8400 is
designed to facilitate use of the priority sample feature. If you plan to use
priority samples during automation, arrange your regular samples in
positions 1-99. Position 0 can be reserved as the Priority Sample position
and will require a minimum of intervention to run. Of course, any vial
position may be designated as a Priority Sample. The Priority Sample
screen has two pages with the following choices.
Page 1
Priority Sample Vial Number
Priority Sample Method
Injection 1 Volume
Injection 2 Volume
34
0 to 99
Method 1 to Method 8
0.1-5, 0.1-10 µL, 1-100 µL
0.1-5, 0.1-10 µL, 1-100 µL
default = 0
default = Method 1
default = 1.0
default = 1.0
03-914852-00:4