Injector Calibration - Agilent Technologies Varian CP-8400 Operator's Manual

Autoinjector
Table of Contents

Advertisement

Setup and Calibration of the CP-8400/8410

Injector Calibration

ACTIVE
METHOD
Start Calib.
To calibrate the Injector 1 and Injector 2 Positions, perform the following
steps:
1. Press the Start Calib. softkey. The CP-8400/8410 will rotate the
2. Using the cursor keys at the right top corner of the CP-3800
3. Scroll to the syringe setting and adjust the height of the needle
CP-8400 AutoSampler & CP-8410 AutoInjector
EDIT
METHOD
CP-8400 INJ 1 CALIB
Component
Direction 1
Tower
CCW
Syringe
Up
Move
tower so that it is over the Injector 1 position and lower the
syringe sled so that it is over the injector. If the needle guide
descends so that it is either resting on the inject switch or below
the needle cone on the inject switch, use the scroll key to
highlight the Syringe up and move the sled up by highlighting the
Up selection and pressing the Move softkey until the needle
guide is just above the injector cone.
keypad, highlight either the CCW (rotates the tower
counterclockwise 10 steps) or CW (rotates the tower clockwise 1
step). Position the tower directly over the injector Inject Switch
Locator cone.
guide such that the sled is just resting on or very slightly
depressing the Injector Switch and is covering the locator cone.
END
RUN
CTRL
TIME
TIME
Direction 2
CW
Down
Finish Calib.
21