Autoinjector
Table of Contents

Advertisement

Installing the CP-8400/8410
Screw Holes: Used
in mounting both
front and rear
AutoSampler
positions.
Screw hole used in
mounting rear position
only.
Carrousel Removed to Show Mounting Screw Locations
CP-8400 AutoSampler & CP-8410 AutoInjector
8. Plug the D-shell connector of the power/communications cable
into the CP-8400/8410.
9. Remove protective shipping materials from the injector carriage
and install the syringe into the AutoSampler injector tower by
opening the clear plastic door, carefully inserting the syringe into
its holder and the needle through its guide, then fasten the
plunger screw on the top and snap the retaining clip over the
syringe.
Screw hole used in
mounting front position
only.
7