Agilent Technologies Varian CP-8400 Operator's Manual page 15

Autoinjector
Table of Contents

Advertisement

Calibrate both the sampling and injector positions of the CP-8400/8410
the first time you install it. Calibrate the injector positions any time you
move the CP-8400/8410 from one position to the other (for instance from
the front mounting position to the rear position). If calibration fails on
either the vial 0 or the injector, you must recalibrate both the injector and
vial 0.
Important Note: Switch the CP-8400/8410 power on before powering up
the CP-3800. Do not turn off the power switch during normal operations.
If you turn the CP-8400/8410 off while the GC is on, or if there is a
momentary communication failure, a communication error fault will occur.
The CP-3800 must be turned off and then on to clear this fault.
CP-3800 Keyboard
To enter Setup, press the Setup key.
ACTIVE
METHOD
[ 1 ] View Instrument Setup
[ 2 ] Edit Instrument Setup
Setup Screen Select item 2, Edit Setup, by pressing 2 on the CP-3800 keypad
and then pressing the Enter key.
12
EDIT
CTRL
METHOD
INSTRUMENT SETUP
END
RUN
TIME
TIME
03-914852-00:4