Operation
Reset Plunger Stroke Counter (Softkey) – Selecting this softkey
causes the Plunger stroke count to be reset to 0.
Injector 1 Needle Depth (%)
Injector 1 Injection Speed (cm/sec)
Injector 2 Needle Depth (%)
Injector 2 Injection Speed (cm/sec)
The Injector Needle Depth parameters define how far into the designated
injector the needle will penetrate. 100% causes the needle to be
completely inserted to its 2-inch length.
The Injector Injection Speed parameters define the rate at which the
needle will penetrate the injector septum and descend to the set Injector
Needle Depth.
CP-8400/8410 Status
The Automation Status page displays the current state of automation and
the CP-8400/8410 progress through the selected automation table.
SPME STATUS
CP-8400 AutoSampler & CP-8410 AutoInjector
0 to 100
default=100
1 to 30
default=30
0 to 100
default=100
1 to 30
default=30
55