Agilent Technologies Varian CP-8400 Operator's Manual page 52

Autoinjector
Table of Contents

Advertisement

Operation
Page 1 Parameters and Softkeys
Sample Penetration Depth (0-100%) – This parameter defines how far
the needle penetrates through the septum into the sample vial. The
percentage refers to the distance from the bottom of the vial septum to
the bottom of the vial interior. The valid range is settable in 1% steps.
The default is 90%.
Solvent Penetration Depth (0-100%) – This parameter defines how far
into the solvent vials the needle will penetrate. The percentage refers to
the distance from the bottom of the vial septum to the bottom of the clean
vial interior. There is only one penetration depth for all three
clean/Internal Standard vials. This parameter is settable in 1% steps. The
default is 90%.
Clean Mode
following parameters:
Number of Pre-Injection Solvent Clean Flushes (0-99) – This
parameter defines the number of times the syringe will draw up solvent,
rotate to the waste cup and expel the syringe contents before the
sampling process begins. The default value is 3.
Number of Post-Injection Solvent Clean Flushes (0-99) – This
parameter defines the number of times the syringe will draw up solvent,
rotate to the waste cup and expel the syringe contents after the injection
process has ended. The default value is 1.
Number of Pre-Injection Sample Clean Flushes (0-99) – This
parameter defines the number of times the syringe will draw up sample,
rotate to the waste cup and expel the syringe contents before the
sampling process begins. The default value is 0.
Clean Solvent Source Vial (I, II, III, I&II, I&III, II&III, I&II&III) – There are
three clean vials. This parameter selects which clean solvent source or
combination of clean solvent sources the AutoSampler will use for its
clean flushes. If a combination Clean Solvent Source is specified, the
system will perform the number of solvent clean flushes as set by the
operator from each of the solvent sources. Consequently, if the operator
selected the Clean Solvent Source of I&II&III and the number of pre-
injection solvent clean flushes was set to three, the system would draw
up and expel solvent from Clean Solvent Source I three times, Clean
Solvent Source II three times and finally Clean Solvent Source III three
times for a total of nine syringe actions prior to beginning the sampling
process. The default value for the Clean Solvent Source is I.
CP-8400 AutoSampler & CP-8410 AutoInjector
Selecting this softkey will display a page with the
49