In order to accurately calibrate vial positions, a calibration tool (CP740889) is used.
The calibration tool simulates a standard vial. The tool is used for
both the CP-8400 and CP-8410 AutoSamplers.
The calibration tool can be placed (depending of the tray-material)
in two different ways in the tray.
See the table below for the correct position.
Calibration Vial in Position 0
Metal tray
Polymer tray
CP-8410 (polymer tray)
calibration tool in vial position 1
For the correct calibration procedure of the sampler vial position, please refer to the
Setup and Calibration section of this manual.
CP-8400 AutoSampler & CP-8410 AutoInjector

Vial calibration tool

CP-8400
Two O-rings visible
-
One O-ring visible
Vial Calibration tool
CP-8410
Calibration Vial in Position 1
Two O-rings visible
One O-ring visible
CP-8400 (metal tray)
calibration tool in vial position 0
Two O-rings