Plunger Retaining Nut
(unscrew to remove
syringe and tighten
when installing)
Syringe Retaining Clip
(lift up to remove
syringe)
Syringe Installation
If you re-locate the CP-8400/8410 to the rear position, you MUST
CAUTION
recalibrate the injector positions. You will not be automatically
prompted to recalibrate the injector positions. If you fail to
re-calibrate the injector positions, the AutoSampler may miss the
injector, resulting in bent syringe needles
8
10. If the currently installed CP-3800 ROM is version 3.1 or lower,
replace with a version 3.3.5 or higher ROM board. To use the
CP-8400 with the CP-3800 Gas Chromatograph, your CP-3800
must have Version 3.2 or later software installed. The CP-8410
requires version 3.3 or later. To use SPME mode sample
introduction with either the CP-8400 or CP-8410 you must have
Version 3.3.5 or later software installed.
11. Replace the left side top cover.
12. Place a vial with cap into the Vial 0 location if you have a
CP-8400 or the Vial 1 location if you have a CP-8410.
13. Place the CP-8400 carrousel onto the base or place the
CP-8410 sample tray oriented so that Vial 1 is under the arc
traced by the tower. Follow the setup procedure described in the
next section to properly calibrate the AutoSampler/AutoInjector
sampling and inject positions.