Neat Sampling Mode

Internal Standard is not permitted with Neat Sampling mode.
Neat mode parameters are set for samples that are normally liquids that
will be injected without dilution.
The Neat sampling mode performs the following sequence of actions:
1. Any pre-injection Clean actions will be executed.
2. A full syringe volume of sample is withdrawn (2 µL/sec) and
3. A full syringe volume of sample is withdrawn (2 µL/sec) and
4. The sample volume is withdrawn (2 µL/sec) and a 1 µL lower air
5. The volume specified in the SampleList is injected. The post
38
expelled (50 µL/sec) into the waste cup. This is performed a total
of six times.
expelled (50 µL/sec) into the sample vial. This is performed three
times.
gap is formed.
injection clean is performed.
03-914852-00:4