16.3 Appendix C –Footswitch SmartLead Sensitivity
Set-Up:
Click on 'Setup', next to the Footswitch sensor inside of the Ultium Hardware Setup (if the Footswitch is
not in this set-up: redetect sensors as described in. The numerical values correspond to the level of
pressure (threshold) needed to activate the Footswitch's internal pressure sensors.
Have the individual wearing the insole stand upright with their feet shoulder-width apart and pointing
straight ahead while parallel to one another. If all four sensors within the insole are activated (heel, M5,
M1, and toe) – the indicator light on the Ultium sensor in which the SmartLead is connected will turn
green. To have the greatest level of sensitivity, use the lowest number in which the indicator light
remains green.
Troubleshooting: If the indicator light is never green, double check that you are using the correct sized
insole (should sit flat on the bottom of the shoe, covering the whole surface, with no overlap of the heel
portion with the toe portion). Repositioning will likely solve this issue.
33
(Rev C May 2019)