Ultium Footswitch SmartLead User Manual
15 Specifications of the Product
15.1 Expected Useful Lifetime
The Footswitch SmartLead has a usable life of seven years.
15.2 Technical Specifications
SmartLead
•
4-zone resistive input with On/Off output.
•
Supports both individual FSR (Force Sensitive Resistors) and insoles.
•
Sensitivity setting to support patients with different body weights.
•
Binary-weighted output voltages for each sensor location are summed for final output voltage.
•
Sample Rate: 2000/4000Hz
•
FSR Sensor response time ~2ms
•
Output Signal Range: 0 to +5V
•
FSR Amplifier Detection Modes: 16 output levels (uniquely detects each sensor combination)
Output voltages for MR3 and analog outputs:
•
No pressure: 0V
•
Heel: 0.266V
•
M5: 0.533V
•
M1; 1.066V
•
Toe: 2.133V
•
All: 4.000V
Physical
•
SmartLead Dimensions:
1.9 cm Width x 1.1 cm Thickness, Length varies slightly
o
SmartLead Weight: 0.2 oz (5.7 g)
15.3 Environmental Conditions for Storage and Transport
•
Ambient Temperature: -40C to +70C
•
Relative Humidity: 10% to 100%
•
Atmospheric Pressure: 500hPa to 1060hPa
(Rev C May 2019)
29