8 Pre-Use Check

8.1 Normal Appearance of Signals

The sensor's STATUS LED provides a means of communicating its operational state. In the idle state,
the STATUS LED will flash blue at a low, once per second rate. When the sensor is actively measuring a
signal, the STATUS LED will flash recognizably faster (green).
Quick Testing:
To ensure that the Footswitch SmartLead is measuring properly when connected to the Ultium Sensor,
firmly press on the center of each of the four regions (heel, 5
shown below:
In the example below, the heel (A), 5
in order. Ensure that the voltage outputs are as follows:
Heel - 0.266V
M5 - 0.533V
M1 - 1.066V
Toe - 2.133V
(Rev C May 2019)
th
metatarsal, 1
th
metatarsal (B), 1
st
metatarsal (C) and toe (D) regions are pressed
st
metatarsal, and toe)