Noraxon Ultium EMG User Manual page 16

Footswitch smartlead
Hide thumbs Also See for Ultium EMG:
Table of Contents

Advertisement

Ultium Footswitch SmartLead User Manual
Troubleshooting: If the indicator light is never green, double check
that you are using the correct sized insole (should sit flat on the
bottom of the shoe, covering the whole surface, with no overlap of
the heel portion with the toe portion). Repositioning will likely solve
this issue.
(Rev C May 2019)
Step 2b
Click on 'Setup', next to the
Footswitch inside of Hardware
Setup. The numerical values
correspond to the level of
pressure (threshold) needed to
activate the Footswitch's
internal pressure sensors.
Have the individual wearing the
insole stand upright with their
feet shoulder-width apart and
pointing straight ahead. If all
four sensors within the insole
are activated (heel, M5, M1,
and toe) – the indicator light on
the Ultium sensor in which the
SmartLead is connected will
turn green. To have the
greatest level of sensitivity, use
the lowest number in which the
indicator light remains green.
Step 3
Once back in the Home screen,
choose to create a new or edit
an existing configuration.
Step 4
In the configuration setup
screen, insert the Ultium
system into the Devices
section.
It is recommended that you
redetect sensors in the
hardware configuration every
time the SmartLeads are
removed from the Ultium