Measurement
Result Example
Spot Measurements
A program example of spot measurements is shown in Table 2-2. This example is
used to apply voltage and measure current.
Spot measurements can be performed by using the following commands.
Function
Selects source function
Sets source output range
Sets source output value
Sets measurement function
Sets measurement range
Sets aperture time in seconds
or by using NPLC value
Enables/disables Ammeter
input
Enables/disables Voltage
Source output
Initiates measurement and
reads result data (latest data)
Result: +3.034160E-02
Keysight B2980 Programming Guide, Ed ition 1
Command
:SOUR:FUNC:MODE VOLT
:SOUR:VOLT:RANG value
:SOUR:VOLT value
[:SENS:]FUNC "func"[, "func"[, "func"]]
[:SENS:]func:RANG:AUTO <ON | OFF>
[:SENS:]func:RANG value
[:SENS:]func:APER time
[:SENS:]func:NPLC value
:INP <ON | OFF>
:OUTP <ON | OFF>
:MEAS?
:MEAS:func?
Programming Examples
Spot Measurements
2- 7