L-770HP--E
8.6 Analog Output Function
This equipment has the two analog outputs and can output the measurement values by the current output
between mA and 0 mA. When an assignment is changed, please set the higher limit value, the lower
limit value, the low cut value, and the damping value to the optimal values again.
Parameter
Assign
Analog output assign
Mode
Output valid / invalid setting
Setting value with the analog output
URV
of 0mA At the setting value, 0mA
is output.
Setting value with the analog output
LRV
of mA At the setting value, mA is
output.
Lowcut
Analog output low cut
Added Damp
Added damping of analog output
Analog output status upon
Error output
the occurrence of an error
8.7 Status Input Function
This equipment can assign the status input or the status output function to the STATUS I/O terminal.
Switching between status input and status output can be done by changing the parameter (Status IO
Assign) setting.
Parameter
Assigning of status input / output
Status IO Assign
function
Input function
Selecting of status input
Input logic
Status input logic
■Status input setting method
()Set Status IO Assign to"Status input" .
()Select the function to assign to the status input by using Input function.
・0% Signal Lock...Stop pulse output and analog output.
・Auto Zero...Perform zeroing.
・Reset Total ...Reset total counter.
()When the STATUS I/O terminal is turned on and kept on for 0. seconds or longer, function is
activated.
(Example) When Input logic is Normal Open, the function is after the terminal is (Close)
for 0. seconds or longer.
Descriptions
Descriptions
Setting range
Mass Flow / Volume Flow
Density / Temp
Drive Output
Normal (valid) / Stop (invalid)
―
―
0 to 0 [%]
0 to 00 [sec]
Downscale:.mA
Zero:mA
Hold:Keep the output.
Upscale:.mA
Setting range
Status input / Status output
0% Signal Lock / Auto Zero
Reset Total / No Function
Normal Open / Normal Close
Standard value
Mass Flow
Normal
―
―
0 %
0 sec
Zero
Standard value
Status input
No Function
Normal Open