Short instructions for use 9 Troubleshooting / warning signals

9.2 Troubleshooting

Malfunction
Nothing works.
The patient chair
does not move.
Display without indic-
ator.
Operating device no
function.
Turbine making loud
running noises.
Satelec Mini LED /
KaVo Poly One does
not work.
No cold light on the
handpieces.
Cause
Main switch is off.
Main service fuse in-
terrupted the electric
circuit.
The safety shutoff is
activated.
(The LED on the con-
trol panel flashes.)
Bus / hardware error. ▶ Turn the device off and on.
Bus / hardware error. ▶ Turn the device off and on.
Turbine wheel faulty. ▶ Replace turbine wheels.
Also refer to: In-
structions for Use of
the Satelec Mini
LED / KaVo Poly One
Cold light not
preselected.
The high-pressure
lamp or Multi LED on
the handpiece is de-
fective.
Remedy
▶ Turn on the main switch.
▶ Unplug the unit from the mains.
▶ Check and replace, if required, the main
service fuse.
The main service fuse is situated next
to the master switch.
▶ For this purpose, open the bayonet
closure with a screwdriver and replace
the fine-wire fuse (T 6.3 H Mat. no.
0.223.2783).
▶ The re-close the bayonet closure with
the screwdriver.
▶ Check the safety shutoff and eliminate
the reason for the shutoff.
▶ Call the service technician to look into
the problem if it continues to exist.
▶ Call the service technician to look into
the problem if it continues to exist.
Follow the operating instructions for the
turbine.
▶ Preselect cold light.
▶ Replace the high-pressure lamp or Multi
LED.
Also refer to: Instructions for Use of
the handpiece
See also:
2 Instructions for Use of the handpiece
28