5� The remote control can be turned off by selecting and confirming "RC Power off"�
In order to change the settings of another MyMode, you first have to switch to that MyMode (see the
section MyMode Switching via remote control)�
INFORMATION
The prosthetist's setting is usually found in the middle of the scale� After adjustment, you can reset
the prosthetist's setting by moving the slider control into the middle�
INFORMATION
The knee joint settings should be optimised by the prosthetist with the help of the X-Soft configura-
tion software� The remote control is not intended for use by the prosthetist in order to set up the
knee joint� The behaviour of the prosthesis can be changed to a certain extent on a day-to-day
basis using the remote control (e�g� while becoming accustomed to the prosthesis)� At the next
appointment, the prosthetist can track the changes in the X-Soft configuration software�
The table below shows the adjustment parameters for the individual MyModes along with a brief
description, the adjustment range of the remote control around the value selected by the prosthetist
and the absolute value range�

Basic mode

Slider name in X-Soft
Stance flexion resistance
Swing phase -
flexion angle
Pitch
Volume
MyModes except friction brake mode
Slider name in
X-Soft
Ascent
Pitch
Volume
Genium X3 - Bionic Prosthetic System
Absolute Range
Remote control
Adjustment Range
120 – 180
±10
55° – 70°
±3°
1000 Hz –
1000 Hz –
4000 Hz
4000 Hz
0 – 4
0 – 4
Absolute Range
Remote control
Adjustment Range
0 – 100
±10
1000 Hz –
1000 Hz –
4000 Hz
4000 Hz
0 – 4
0 – 4
Description
Resistance to flexion movements, e�g�
while walking downstairs or sitting
down in a chair�
Maximum flexion angle in the swing
phase�
Frequency (pitch) of the beeper for
confirmation signals�
Volume of the beeper for confirmation
signals�
Description
Value for speed at which flexion resist-
ance increases as the knee angle
increases�
Frequency (pitch) of the beeper for
confirmation signals�
Volume of the beeper for confirmation
signals�