Otto Bock Genium X3 Series Patient Information page 23

Bionic prosthetic system
Hide thumbs Also See for Genium X3 Series:
Table of Contents

Advertisement

Type of error
Event
Minor error
e�g� service interval has
expired
e�g� a sensor (tube adapter,
triple sensor, ���) is not
operating
Medium error
Selftest failure (if signals
occur after removing the
charger)
AXON Tube Adapter has
been disconnected while the
knee joint is activated and
the inductive charger is not
AXON
attached
Tube Adapter
AXON Tube Adapter not
disconnected
connected while removing
inductive charger
e�g� failure of the valve drive
Selftest failure (if signals
Major error
occur after removing the
charger)
Total failure
No electronic functionality
Charging Sta-
tus 0%
Error code on
the display of
the Remote
Control
As soon as the knee joint identifies an error in the knee joint, at least 5 beeps and 5 slow pulsating
vibration signals are emitted� The knee joint activates a safety mode and remains in this state until
the error is corrected (as long as the battery capacity is sufficient)�
In addition to basic mode and the MyModes, the knee joint has various safety modes that offer dif-
ferent levels of residual functionality depending on the type and severity of the error�
The OPG function, stairs and obstacles function, Intuitive Stance function and Sitting function are
deactivated when the joint is in safety mode� Beyond that, the following levels of residual functional-
ity are possible:
• Continuous high flexion resistance
• Continuous stance phase flexion resistance with the ability to initiate the swing phase
Depending on the type of error, protection against swing phase initiation while walking backwards
may or may not be active� You must therefore exercise caution when walking backwards after an error�
Genium X3 - Bionic Prosthetic System
User restriction
No restriction
Walking possible with
restrictions (activation
of safety mode)
High stance resis-
tance
Swing phase no lon-
ger possible
Activation of safety
mode (dependant on
the remaining func-
tionality)
Swing phase no lon-
ger possible
Required action
Consult Prosthetist
Restart the knee joint by
attaching and removing in-
ductive charger (up to three
times); If error remains, con-
sult Prosthetist immediately
Connect the AXON Tube
Adapter
Restart the knee joint by
attaching and removing in-
ductive charger (up to three
times); If error remains, con-
sult Prosthetist immediately
Consult Prosthetist im-
mediately
Charging required
Consult Prosthetist
immediately(Error Code
number required)
Ottobock | 23