Connecting
4.2 Electrical connections
4.2.2.1
Connecting the power supply
Procedure
Note
Appliance plug
You can use any appliance plug that fulfills the conditions of DIN EN 60320-1, ≥6 A.
The power supply cable must comply with the valid regulations and conditions for the place of
installation and be provided with a protective conductor at the potential of the housing. The
cross-section of each wire must be ≥ 1 mm
● Connect the phase conductor to the marked position in the plug.
● Make sure that a power disconnection device is located in the direct vicinity of the device
(for load rating, see nameplate).
● Make sure that the supply voltage matches the voltage specified on the nameplate of the
device.
①
②
③
Figure 4-1
42
5
4
Power supply unit
Equipotential bonding
Interlock for Ethernet plug
Connecting the power supply
.
2
3
2
1
④
Appliance plug
⑤
Securing bracket for power connector
Compact Operating Instructions, 05/2015, A5E35134047-02
Devices in explosion-proof design