Leak Testing For Oxymat 7 - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Commissioning
6.3 Leak testing of containment system
Recommended test setup
To check for leaks, we recommend the test setup shown below:
Figure 6-2
6.3.3

Leak testing for OXYMAT 7

Procedure
1. Close the reference gas inlet.
2. Connect the sample gas outlet to a pressure measuring instrument.
3. Apply a pressure of approximately 3 000 hPa (absolute pressure) to the measuring inlet.
Then block the measuring inlet.
4. Wait for approximately 5 minutes until the sample gas that flowed in has reached the
ambient temperature and the pressure is stable. Note the pressure.
5. Wait an additional 5 minutes and note the pressure again.
If the pressure change is less than 10 hPa, the sample gas path is adequately tight.
68
Blanking plug (on gas connection 4)
Shut-off valve
OXYMAT7: Recommended test setup to check for leaks
Pressure regulator
Manometer (gauge pressure)
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02