Siemens SIPROCESS GA700 Compact Operating Instructions page 177

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

5. Define measuring ranges:
[2.03] Settings > Measuring ranges
Note
Putting back into operation with disabled measuring ranges
The measuring ranges are activated in the factory. If you disable the measuring ranges
(setting: "Off"), all the setting options in other menus related to the measuring range are
also hidden. The measuring ranges can no longer be used.
If you put the device back into operation with fully deactivated measuring ranges, the status
bar of the display shows message 150 "Parameter value invalid". Before you continue to
put the device back into operation:
● Activate at least one measuring range.
● Disable the autoranging function if it is enabled.
6. Set inputs/outputs: [2.09] Inputs/outputs
– Set digital inputs: [2.09.3] Digital inputs
– Setting digital outputs [2.09.5] Digital outputs
If you have the appropriate optional modules:
– Set analog inputs: [2.09.1] Setting analog inputs
– Set analog outputs: [2.09.2] Setting analog outputs
7. Check setting of gas path and change, if necessary:
[2.12] Settings > Gas path/process tag label
8. Calibrate the device: [2.08] Settings > Calibrations
– Familiarize yourself with the basics of calibration. Calibration changes internal device
– Prepare for the calibration. Set the setpoints, calibration tolerances, and the sample gas
– Assign parameters for a calibration process: [2.08.3] Execute calibrations
9. Save parameter set: [3.10] Maintenance & Diagnostics > Save/load parameter set
Approach in case of an incomplete/faulty boot process
If the device detects a serious error during booting, the boot process is aborted. The start
screen is no longer updated. In this case, contact the service technician responsible.
See also
References (Page 91)
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02
characteristics.
purging time: [2.08.1] Setpoints, tolerances, purging time
6.4 Commissioning the SIPROCESS GA700
Commissioning
71