Cable Glands For The Wall-Mounted Device; Rack-Mounted Device; Connecting The Power Supply - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting
5.1 Electrical connections
5.1.2.3

Cable glands for the wall-mounted device

NOTICE
Cable glands for the wall-mounted device
Match the diameter of the cable to the threaded cable glands. After installation, make sure
that the cables are secure and that the seals are in the correct position.
5.1.3

Rack-mounted device

5.1.3.1

Connecting the power supply

Requirements
Note
Appliance plug
An appliance plug is included in the scope of delivery. However, you can use any appliance
plug which meets the requirements of DIN EN 60320-1, ≥6 A.
Note
Connection information
The power cable must comply with the valid regulations and conditions for the place of
installation and be provided with a protective conductor at the potential of the enclosure. The
cross-section of each wire must be ≥ 1 mm
● Connect the phase conductor, neutral conductor (N) and the protective conductor (PE) to
the designated position in the plug.
● Make sure that a power disconnection device is located in the direct vicinity of the device
(for load rating, see nameplate).
● Make sure that the supply voltage matches the voltage specified on the nameplate of the
device.
44
.
2
Compact Operating Instructions, 05/2015, A5E31805656-03
Quick Start